Controlling Acquisition
Operation
340
Select Acquisition Mode
Press the main menu button Mode. Use the
side menu to select from these modes.
H Press Sample to select sample mode.
H Press Peak Detect to select peak detect mode. Peak detect mode only
operates at SEC/DIV settings of 25 ms or slower.
H Press Envelope to select envelope mode. Use the General Purpose
Knob to adjust the number of envelopes (2, 4, 8, 16, 32, 64, 128, 256, or
R).
H Press Average to select average mode. Use the General Purpose Knob
to adjust the number of acquisitions to average (2, 4, 8, 16, 32, 64, 128,
or 256).
Activate Single Sequence Mode
Press the main menu button Stop
After and then the side menu button Single Acquisition Sequence.
When in single sequence mode, the oscilloscope acquires a single se
quence, display it, and then waits for you to press the RUN/STOP button.
The exact nature of a sequence" depends on the present acquisition mode.
In sample or peak detect mode, a sequence" is a single acquisition. In
envelope or average mode, a sequence" is a number of acquisitions equal
to the number of envelopes or averages.
For example, when both average and single sequence mode are active and
the number of averages is 16, the oscilloscope collects 16 acquisitions,
averages them, displays the resulting waveform, and stops until you press
the RUN/STOP button.
To deactivate single sequence mode, press the side menu button RUN/STOP