88 / 242 page

Taking Measurements
Operation
336
Cursor
Figure 319:The CURSOR Button
Figure 320:The Cursor Menu
2. If you want to take a time measurement in Hertz instead of seconds,
press the main menu button Time Units and use the side menu to switch
between the two.
3. If the main menu selection Function is not highlighted, press its main
menu button.
4. Use the side menu to select the type of measurement you want to take.
Select H Bars to take a voltage measurement, select V Bars to take a
time measurement, or select Paired to take simultaneous voltage and
time measurements.