89 / 242 page

Taking Measurements
TDS 310, TDS 320 & TDS 350 Instruction Manual
337
5. Use the General Purpose Knob to move one cursor to a desired mea
surement point. For example, if you wanted to measure both the peak to
peak voltage and period of a sine wave, you would start by moving one
paired cursor to a trough" (see Figure 321).
Figure 321:Paired Cursor Measurements of a Sine Wave
6. Press TOGGLE.
7. Use the General Purpose Knob to move the second cursor to another
desired measurement point. To continue the example, you would move
the second cursor to the previous peak."
8. Read the measured value(s) at the upper right corner of the display next
to the D sign(s). Figure 321 shows a sine wave with a 5.48 Vpp and a
13.2 ms period (6.6 ms X 2).