| поискавой системы для электроныых деталей |
|
MAXQ3180-RAN+ датащи(PDF) 45 Page - Maxim Integrated Products |
|
|
|||||||||||||||||||||||||||||
MAXQ3180-RAN+ датащи(HTML) 45 Page - Maxim Integrated Products |
|
45 / 48 page ![]() 3) Measure average raw VRMS by reading the RAM register several times. 4) Measure average NS by reading the RAM register several times. 5) Calculate coefficient K = [NS/216] x [VRMS/216]/216. 6) Calculate Eoff_hi coefficient: Eoff_hi = [(P2 x I1 - P1 x I2)/(I1 - I2)])/K. 7) Calculate Eoff_lo coefficient: Eoff_lo = [(P4 x I3 - P3 x I4)/(I3 - I4)]/K. 8) Calculate Gain_lo coefficient (using calculated above Eoff_hi, Eoff_lo, and K): Gain_lo = ({(I3/I1) x (P1 + K x Eoff_hi)/ (P3 + K x Eoff_lo)} – 1) x 216 Note: All intermediate calculations should be per- formed with double precision, the last result rounded to the nearest integer. Eoff_hi, Gain_lo, and Eoff_lo are 16- bit signed coefficients. If a calculated value exceeds 16 bits, the part cannot be calibrated. In this case, assign the largest possible value, i.e., 0x7FFF for positive over- flow or 0x8000 for negative overflow. Phase Angle Calibration This calibration is intended to compensate for phase angle errors across the range of input loads. The MAXQ3180 uses four PA_n coefficients to calculate the compensation phase angle as a 3rd-order polynomial: compensation_phase_angle = PA_0 + (PA_1 x X) + (PA_2 x X2) + (PA_3 x X3) Where X is proportional to the logarithm of raw IRMS. Full calibration of all four coefficients requires at least four measurements of the phase angle across the range. However, in most cases, only one coefficient PA_0 is needed that requires at least one phase angle measurement. For input signals, all three phase nominal voltages 220V should be applied. Only one current signal is applied to the phase input being calibrated. Other cur- rent inputs should be 0. Phase Angle Error Measurement This calibration procedure requires measuring the phase angle error of the meter’s output at the given current input level I_in. This can be done in two ways: (I) by reading raw RAM registers, or (II) by measuring pulse output errors at different power factors. Measurement Technique I: Measure Output Phase Angle Error by Reading RAM Registers 1) Measure average raw active power by reading the RAM registers several times. 2) Measure average raw reactive power by reading the RAM registers several times. 3) Calculate output phase angle: output_PA = atan(Reactive_raw_pwr/ Active_raw_pwr) 4) Calculate output phase angle error: output_PA_error = output_PA - input_PA Measurement Technique II: Measure Output Phase Angle Using Pulse Output Before the pulse output can be measured, the pulse configuration registers PLSCFG and THR1 (or THR2) must be properly configured. The pulse output should be set to active power. For details on how to set those registers, see the Power Gain Calibration section. 1) Apply input signal with power factor 1 (input phase angle 0 °) to the meter’s inputs. Measure pulse output error for active power, in percent. Denote it as Err00 for future reference. 2) Apply input signal with power factor 0.5L (input phase angle 60 °) to the meter’s inputs. Measure pulse output error for active power, in percent. Denote it as Err60 for future reference. 3) Calculate output phase angle error: output_PA_error = atan{(1/sqrt(3)) x (Err00 – Err60)/(100% + Err00)} Phase Angle Calibration Procedure 1) Clear PA_0, PA_1, PA_2, PA_3 coefficients being calibrated to 0x0000. 2) Phase angle error measurement. a) Apply current input signal of the RMS value close to IFS/21.5 to the phase input being calibrated. For example, typical target value is 100/21.5 = 35.36A, so reasonable input current is 30A or 40A. Measure the output phase angle error as described above. Denote applied current as I1 and measured phase angle error expressed in radians as PAerr1 for future reference. b) Measure average raw IRMS by reading the RAM register several times. Denote this as I1_raw for future reference. Low-Power, Multifunction, Polyphase AFE ______________________________________________________________________________________ 45 |
|
ссылки URL |
| Вашему бизинису помогли Аллдатащит? [ DONATE ] |
Что такое Аллдатащит | реклама | контакт | Конфиденциальность | Ссылка на техническое описание | обмен ссыками | поиск по производителю All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |